Kyros Engineering Engineering Tools
Home › Tools › Sensor Signal Conditioner
// Sensor tools Proto v0.1ThermistorRTDThermocoupleStrain gauge

Sensor Signal Conditioner

Pick a sensor, get its signal and where it goes wrong. Thermistors, RTDs, thermocouples, bridges and strain gauges each have a tab, and the last tab follows the signal through the amplifier, filter and ADC with a noise and error budget.

Runs entirely in your browser. Nothing is uploaded.

1 · NTC thermistor

Ω
K
Steinhart-Hart coefficients (optional)

Fill all three to override Beta. Blank uses the Beta model.

Fit Steinhart-Hart from three datasheet points
°C
Ω
°C
Ω
°C
Ω
—

2 · Voltage divider

V
Ω

3 · Linearization and readback

°C
°C
Ω
Temperature
—

Result

Resistance vs temperature

Divider output vs temperature

Linearization

MethodSeries resistorPeak nonlinearity

Peak nonlinearity is the largest gap between the divider output and a straight line through the two range ends, expressed in °C.

R-T table

T (°C)R (Ω)Vout (V)

1 · RTD

Ω
Ω
%
mA
mW/°C

2 · Callendar-Van Dusen (IEC 60751)

R(T) = R0 (1 + A T + B T²)                T ≥ 0 °C
R(T) = R0 (1 + A T + B T² + C (T − 100) T³)   T < 0 °C
A = 3.9083e-3   B = −5.775e-7   C = −4.183e-12
α = 0.00385 Ω/Ω/°C

3 · Readback

Ω
Temperature
—

Result

Resistance vs temperature

Measurement error

Lead-wire error and self-heating, in °C of reading error.

R-T table

T (°C)R (Ω)V (mV)Wire error (°C)Self-heat (°C)

IEC 60751 tolerance classes

Tolerance of the element itself, ± °C: AA 0.1 + 0.0017|t|, A 0.15 + 0.002|t|, B 0.3 + 0.005|t|, C 0.6 + 0.01|t|.

1 · Thermocouple type

2 · Temperature → voltage

°C
°C

3 · Voltage → temperature

mV
°C
°C

Result

Voltage vs temperature

NIST ITS-90 reference function against a 0 °C cold junction.

Sensitivity (Seebeck coefficient)

1 · Bridge

     +Vex
    ┌──┴──┐
   R1     R3
    ├─ A ─┼─ B ─ → Vout = V(B) − V(A)
   R2     R4 (sensor)
    └──┬──┘
      GND
Ω
Ω
Ω
Ω
Ω
µε
µε
V
V

Result

Bridge output vs R4 change

Sensitivity

Signal chain

→ → → →

Sensor

mV
mV
Ω
%
µV rms

Bridge / divider

V
µV

Amplifier

×
kHz
nV/√Hz
dB
µV
%

Filter

Hz
µV rms

ADC

V
SPS
LSB

Result

Signal level through the chain

Half the sensor span, in mV, after each stage.

Noise budget

µV rms at the ADC input. Log scale.

Noise by stage

StageSignal (mV)Noise (µV rms)SNR (dB)Share of noise power

Accuracy budget

Static errors referred to the amplifier input.

TermµV at input% of span
// How it works

From a physical quantity to ADC counts

Every sensor here turns temperature or strain into a small change in resistance or voltage. The tool computes that change from the standard reference function for the sensor, then shows what limits it: linearity, lead wires, self-heating, cold-junction error, noise, and converter resolution.

Thermistors

Beta model R(T) = R25 · exp[ B (1/T − 1/298.15) ] α = −B/T² Steinhart-Hart 1/T = A + B ln R + C (ln R)³ (T in kelvin) fit from 3 points: solve the 3×3 linear system for A, B, C divider (NTC at the bottom) Vout = Vref · R / (R + Rs)

The Beta model is a two-point fit and drifts away from the real curve outside the 25/85 °C pair. Steinhart-Hart with three coefficients follows it to a few millidegrees over a 100 °C span. The linearization table compares three series resistors: the geometric mean of the end resistances, the inflection formula Rs = R(Tc)(B − 2Tc)/(B + 2Tc) that puts the divider's inflection point at the centre, and the resistor that minimizes the peak error found by a direct search.

RTDs

T ≥ 0 °C: R = R0 (1 + A T + B T²) T < 0 °C: R = R0 (1 + A T + B T² + C (T − 100) T³) 2-wire error 2 R_lead / (dR/dT) 3-wire: only the lead mismatch remains 4-wire: none self-heating ΔT = I² R / δ largest current for 0.1 °C: I = √(0.1 · δ / R)

The wire error uses the slope of the curve at each temperature, which falls from 0.39 Ω/°C at 0 °C to about 0.32 Ω/°C at 850 °C for a Pt100. The dissipation constant δ is the power that raises the element 1 °C above its surroundings and depends on the package and the medium it sits in. Moving air gives a much smaller value than a stirred liquid.

Thermocouples

E(T) = Σ c_i T^i (+ a0 · exp[a1 (T − a2)²] for type K above 0 °C) NIST ITS-90, mV against 0 °C meter reads E(hot) − E(cold) hot = E⁻¹( reading + E(cold) ) cold-junction error effect ΔT_hot = ΔT_cj · S(cj) / S(hot)

A thermocouple measures a temperature difference. The voltage you read needs the cold-junction temperature added back, as a voltage and not as degrees. The tool inverts the NIST polynomial numerically, so it stays exact over the whole range instead of using the separate approximate inverse polynomials. The last line shows why a 1 °C error at the cold junction becomes nearly a 1 °C error at the hot junction when both sit at similar sensitivities.

Bridges and strain gauges

Vout = Vex · [ R4/(R3 + R4) − R2/(R1 + R2) ] balanced when R1 R4 = R2 R3 quarter bridge Vout/Vex = x/(4 + 2x) (x = GF·ε; ≈ GF·ε/4) half bridge GF·ε/2 full bridge GF·ε

A quarter bridge is slightly nonlinear: at GF = 2 and 1000 µε the exact output is 0.1 % lower than the usual GF·ε/4 formula. Half and full bridges with gauges on opposing arms are linear.

Signal chain

source noise √(4 k T R_s · B) 4.07 nV/√Hz per kΩ at 300 K amplifier noise e_n · √B_eq noise bandwidth of an n-pole Butterworth low-pass = fc · (π/2n) / sin(π/2n) quantization LSB / √12 ideal SNR = 6.02 N + 1.76 dB total root-sum-square of every term at the ADC input; ENOB = (SNR − 1.76) / 6.02

Noise is referred to the ADC input so the terms add directly. A filter limits how much amplifier noise reaches the converter and also stops out-of-band signals aliasing. The tool warns when the filter cutoff is above half the sample rate, when the swing clips the ADC, and when it uses so little of the range that more gain would help.

Assumptions and limits

References

  1. NIST, ITS-90 Thermocouple Database, Standard Reference Database 60 (srdata.nist.gov/its90): reference functions and tables for types B, E, J, K, N, R, S, T. Coefficients were taken from the public-domain table in the Python package thermocouples_reference 0.20 and checked against NIST table values.
  2. IEC 60751:2008, Industrial platinum resistance thermometers and platinum temperature sensors: Callendar-Van Dusen coefficients and tolerance classes.
  3. J. S. Steinhart, S. R. Hart, "Calibration curves for thermistors," Deep-Sea Research 15, 1968, 497–503.
  4. National Instruments, Measuring Strain with Strain Gages, and Hottinger Brüel & Kjær, An Introduction to Measurements using Strain Gages: bridge equations and quarter-bridge nonlinearity.
  5. Analog Devices, Linear Circuit Design Handbook (ed. H. Zumbahlen), chapters on sensors, noise and converters; Texas Instruments application notes on RTD, thermocouple and bridge front ends.
  6. H. Johnson, Thermal agitation of electricity in conductors, Phys. Rev. 32, 1928; H. Nyquist, Phys. Rev. 32, 1928.

Sensor reading off by a degree?

Kyros builds measurement front ends: sensor selection, excitation, cold-junction and lead compensation, low-noise amplification and the firmware that calibrates it in production.

Talk about your project →