Ear Acoustics Explorer

Model the ear canal as an acoustic transmission line ending in a lumped middle-ear network. See how fluid, pressure and ossicle problems change absorbance, the tympanogram and the reflected waveform, then size the probe: sample rate, FFT length, SNR and stimulus band.

Educational / engineering model. Not a medical device. Not for diagnosis. Every number here comes from a simplified model calibrated to published group averages. It says nothing about any real person's ear.
Runs entirely in your browser. Nothing is uploaded.

1 · Condition

Middle ear

0 %
daPa
cm³
mm
mm

Stiffness

×
×
×
daPa

2 · Ear canal & probe

cm
mm
cm
cm

3 · Measurement design

Hz
Hz
Hz
ms
dB SPL
dBA
×
ΔA

Result

Simulated 226 Hz tympanogram
—
Peak Ytm
—
Peak pressure (TPP)
—
Ear-canal volume
—
Tympanometric width
—
Absorbance 1–2 kHz
—
Spectral gradient (approx.)
—
—
MeasureModelNormative 90 % rangeSource
Engineering estimate from a simplified model. Not a medical device and not for diagnosis.

Wideband absorbance

A = 1 − |R|² at the probe. Shaded: normal-adult 90 % range. Dashed: the normal ear for the same age group and canal.

Tympanogram

Admittance magnitude at the probe while the ear-canal pressure sweeps. Shaded: normal peak-pressure region (−100 to +50 daPa).

Incident vs reflected

The stimulus, and the wave that comes back from the eardrum to the probe (inverse FFT of the stimulus spectrum × R(f)). The probe mic actually hears their sum.

Acoustic reflectometry

An unsealed tip at the canal entrance hears incident + reflected sound. Reflection from the eardrum cancels the incident wave near the quarter-wave frequency, c/(4L). A stiffer, fluid-backed eardrum reflects more and deepens the dip. Shaded: the 1.8–4.4 kHz sweep used by commercial reflectometers.

Dip frequency
—
Dip depth
—
Gradient angle
—
The angle uses reference points ±200 Hz from the dip, on axes where 1 kHz is drawn as long as 10 dB. Commercial devices use their own scaling, so this number does not map to a device reading or its cut-offs.

Measurement design

What the electronics have to deliver for this canal and condition. All values are engineering estimates.

Frequency resolution
—
Record length
—
Round trip 2L/c
—
Quarter-wave null at probe
—
SNR available (worst bin)
—
SNR needed for ΔA
—
Minimum stimulus level
—
Recommended band
—
——

Per-bin SNR across the band

Network element values for this condition
BlockElementSICGS
// How it works

A tube, a lumped middle ear, and the numbers a probe has to measure

Every acoustic middle-ear test does the same thing: put a known sound into a sealed (or nearly sealed) ear canal and work out how much of it the eardrum takes in and how much it sends back. Tympanometers do it at one tone while the canal pressure sweeps. Wideband acoustic immittance (WAI) systems do it across 0.2–8 kHz. Acoustic reflectometers, including the smartphone-plus-paper-funnel approach, listen to the interference between incident and reflected sound near the canal's quarter-wave frequency. This page models all three with one physics chain.

1. The ear canal as a transmission line

Below about 8 kHz the canal carries plane waves, so it behaves as a uniform tube of area A = πd²/4 between the probe tip and the eardrum (residual length L = canal length − insertion depth). Air properties are taken at body temperature (37 °C: c = √(γRT) = 353 m/s, ρ = 1.14 kg/m³).

Z0 = ρc / A -- characteristic impedance Zp = Z0 · (Ztm + j·Z0·tan kL) / (Z0 + j·Ztm·tan kL) -- probe impedance, k = ω/c (lossless) R = (Z − Z0) / (Z + Z0) -- pressure reflectance, at the probe or the eardrum A = 1 − |R|² -- absorbance (Feeney et al. 2013 terminology) Rp = Rtm · e^(−2jkL) so |Rp| = |Rtm| -- absorbance does not depend on probe depth f_null = c / (4L) -- pressure null at the probe (Siegel 1994)

With losses switched on, the propagation constant and Z0 pick up first-order viscous and thermal boundary-layer terms (Γ = (jω/c)·√((1+εv)(1+εt)), εv = (1−j)δv/a, εt = (γ−1)(1−j)δv/(a√Pr), δv = √(2μ/ρω)). That adds a little absorbance in the canal itself, so the probe reads slightly more than the eardrum.

2. The middle ear as a lumped network

The eardrum load uses the topology of Kringlebotn's (1988) network for the human middle ear, in a Zwislocki-style analog where series elements share volume velocity:

Z_ME = Z_cav + Z_eardrum + Z_chain -- in parallel with Z_perf when the eardrum has a hole Z_cav = C_tympanum ‖ (R_aditus + jωM_aditus + C_mastoid) Z_eardrum = jωLd + [ (Rs + jωLs + 1/jωCs) ‖ (Rr + 1/jωCr) ] -- plus φ·(R_f + jωM_f) with fluid Z_chain = (Rmc + 1/jωCmc) ‖ ( Ro + jωLo + [ (Ri + 1/jωCi) ‖ (Rc + jωLc + 1/jωCc) ] ) C = V / (ρc²) = V / (γ·P0) -- compliance of an air volume V G_w = 0.228 mmho -- tissue-loss shunt at the eardrum (canal wall, eardrum; pressure-independent)

Starting values are Kringlebotn's, read from S. Lansbergen's MIT-licensed MATLAB implementation of that network. The code here is a new JavaScript implementation of the same topology. The element values were then refitted by least squares, pulled toward the original values, so that the model's normal adult hits:

The fit moved no element more than 2.5× from its original value. The table below is the evidence:

FrequencyModel, normalPublished mean ± SDModel, full effusionPublished effusion mean ± SD
226 Hz measureModel, normal adultMargolis & Heller (1987) mean, 90 % range

3. Static pressure and the tympanogram

A tympanometer sweeps the canal from +200 to −400 daPa. The pressure difference across the eardrum, Δp = p_canal − p_middle-ear, stretches it. That is modelled with an empirical stiffening law applied to the eardrum compliances Cs and Cr. It follows the same idea as Vanhuyse et al. (1975): reactance is a symmetric, peaked function of pressure.

s(Δp) = 1 + (|Δp| / w)^1.2 -- w = 13 daPa (adult), 17 daPa (child preset) Cs → Cs / s, Cr → Cr / s -- peak where p_canal = p_middle-ear, so TPP ≈ middle-ear pressure 1 acoustic mmho = 1e-3 cm⁵/(dyn·s) = 1e-8 m³/(Pa·s) 1 cm³ of air ≈ 1.0 mmho at 226 Hz -- ASHA 1988 ECV = |Y| at +200 daPa (226 Hz) Ytm = |Y|peak − |Y|(+200 daPa) -- positive-tail compensation TW = pressure interval at half Ytm -- de Jonge 1986, via ASHA 1988

w and the exponent were set so that the normal adult reproduces the Margolis & Heller means for Ytm, ECV and TW together. The residual eardrum admittance at +200 daPa makes ECV read about 20 % high, as real tympanometers do (Shanks & Shohet 2009 report 24–39 %). Pressurizing only the eardrum branches also gives the behavior seen in wideband tympanometry: as |Δp| grows, low-frequency absorbance collapses and the absorbance ridge moves up toward 3–5 kHz.

Classification follows Jerger (1970) and Jerger et al. (1972), with numeric limits from the Margolis & Heller 90 % ranges:

At 1000 Hz the canal and eardrum are near resonance, so tympanograms can be notched. The tool shows them but does not type them.

4. Conditions: what each preset changes

5. Acoustic reflectometry and the spectral gradient

Reflectometry was introduced by Teele & Teele (1984). An unsealed probe at the canal entrance sends a 1.8–4.4 kHz sweep and records incident plus reflected sound. Near the quarter-wave frequency the two cancel and the response shows a dip. A mobile eardrum absorbs energy and gives a broad, shallow dip. Fluid behind the eardrum reflects more and gives a narrow, deep one.

The commercial device reduces the dip to a "spectral gradient" angle between the slopes on either side, with cut-offs near 95° and 75° in its own units (Combs et al., US 5,699,809). The smartphone version (Chan et al. 2019, the research behind WavelyDx) plays 150 ms, 1.8–4.4 kHz FMCW chirps through a paper funnel with a 7 mm tip, records at 48 kHz, finds the dip in 2.3–3.8 kHz, and classifies the shape around it. It reported an AUC of 0.898 versus 0.776 for a commercial reflectometer. The angle alone, computed with ±200 Hz reference points, gave an AUC of only 0.687.

Here the response is modelled as

H(f) = 1 + R_eardrum(f) · e^(−2ΓL) L = canal length + tip offset; dip depth ≈ 20·log10(1 − |R|)

The angle uses the same ±200 Hz reference points. Treat it as an approximation: the axis scaling is ours, and the model's funnel is an idealized continuation of the canal.

6. Measurement design

Δf = fs / N, record T = N / fs, round trip τ = 2L / c separation: τ ≥ 2/B and τ·fs ≥ 2 -- 2/B = main lobe of a pulse of bandwidth B N0 = p_ref² · 10^(L_self/10) / 19 980 Hz -- mic noise PSD, A-weighted self-noise taken as flat SNR_k = |X_k|² · M / (N · N0 · fs/2) -- per DFT bin, M averages δA = 2|R|·δ|R| ⇒ SNR ≥ 20·log10(√2 · k · |R| / ΔA) -- k = 2 ≈ 95 % minimum level = level + SNR_needed − SNR_worst-bin

Three results usually surprise people:

Limits

References

  1. Feeney MP, Hunter LL, Kei J, et al. Consensus statement: Eriksholm workshop on wideband absorbance measures of the middle ear. Ear Hear 2013;34 Suppl 1:78S–79S.
  2. Karuppannan A, Barman A, Mamatha NM. Wideband absorbance pattern and its diagnostic value in adults with middle ear effusions and tympanic membrane perforation. J Int Adv Otol 2024;20(2):158–163 (normal and effusion absorbance, Table 2).
  3. Margolis RH, Heller JW. Screening tympanometry: criteria for medical referral. Audiology 1987;26:197–208. Ranges as tabulated in Shanks J, Shohet J, Tympanometry in clinical practice, in Katz J (ed.), Handbook of Clinical Audiology, 6th ed., ch. 9, Table 9.1 (2009).
  4. American Speech-Language-Hearing Association. Tympanometry (ASHA Working Group on Aural Acoustic-Immittance Measurements), 1988: definitions of TW, the 226 Hz volume equivalence.
  5. Jerger J. Clinical experience with impedance audiometry. Arch Otolaryngol 1970;92:311–324. Jerger J, Jerger S, Mauldin L. Studies in impedance audiometry. I. Normal and sensorineural ears. Arch Otolaryngol 1972;96:513–523 (As and Ad subtypes).
  6. Molvaer OI, Vallersnes FM, Kringlebotn M. The size of the middle ear and the mastoid air cell. Acta Otolaryngol 1978;85:24–32.
  7. Wiley TL, Oviatt DL, Block MG. Acoustic-immittance measures in normal ears. J Speech Hear Res 1987;30:161–170 (226 Hz phase 56–76°).
  8. Kringlebotn M. Network model for the human middle ear. Scand Audiol 1988;17(2):75–85.
  9. Lansbergen S. middle_ear_model: lumped-element dynamic middle-ear model with pressure variation (MATLAB). MIT License. github.com/Lansbergen/middle_ear_model. Topology and starting element values; no code copied.
  10. Zwislocki J. Analysis of the middle-ear function. Part I: Input impedance. J Acoust Soc Am 1962;34:1514–1523.
  11. Vanhuyse VJ, Creten WL, Van Camp KJ. On the W-notching of tympanograms. Scand Audiol 1975;4:45–50.
  12. Voss SE, Rosowski JJ, Merchant SN, Peake WT. Middle-ear function with tympanic-membrane perforations. II. A simple model. J Acoust Soc Am 2001;110:1445–1452.
  13. Stepp CE, Voss SE. Acoustics of the human middle-ear air space. J Acoust Soc Am 2005;118:861–871.
  14. Ravicz ME, Rosowski JJ, Merchant SN. Mechanisms of hearing loss resulting from middle-ear fluid. Hear Res 2004;195:103–130.
  15. Ellison JC, Gorga M, Cohn E, Fitzpatrick D, Sanford CA, Keefe DH. Wideband acoustic transfer functions predict middle-ear effusion. Laryngoscope 2012;122:887–894.
  16. Beers AN, Shahnaz N, Westerberg BD, Kozak FK. Wideband reflectance in normal Caucasian and Chinese school-aged children and in children with otitis media with effusion. Ear Hear 2010;31:221–233.
  17. Merchant GR, Al-Salim S, Tempero RM, Fitzpatrick D, Neely ST. Improving the differential diagnosis of otitis media with effusion using wideband acoustic immittance. Ear Hear 2021;42:1183–1194.
  18. Rosowski JJ, Nakajima HH, Hamade MA, et al. Ear-canal reflectance, umbo velocity, and tympanometry in normal-hearing adults. Ear Hear 2012;33:19–34. Nakajima HH, et al. Comparison of ear-canal reflectance and umbo velocity in patients with conductive hearing loss. Ear Hear 2012;33:35–43.
  19. Voss SE, Allen JB. Measurement of acoustic impedance and reflectance in the human ear canal. J Acoust Soc Am 1994;95:372–384. Keefe DH, Ling R, Bulen JC. Method to measure acoustic impedance and reflection coefficient. J Acoust Soc Am 1992;91:470–485. Allen JB. Measurement of eardrum acoustic impedance. In Peripheral Auditory Mechanisms, Springer, 1986.
  20. Siegel JH. Ear-canal standing waves and high-frequency sound calibration using otoacoustic emission probes. J Acoust Soc Am 1994;95:2589–2597.
  21. Teele DW, Teele J. Detection of middle ear effusion by acoustic reflectometry. J Pediatr 1984;104:832–838. Combs JT, et al. Device and process for generating and measuring the shape of an acoustic reflectance curve of an ear. US Patent 5,699,809 (1997).
  22. Chan J, Raju S, Nandakumar R, Bly R, Gollakota S. Detecting middle ear fluid using smartphones. Sci Transl Med 2019;11:eaav1102. Product description: wavelydx.com.
  23. Keefe DH. Acoustical wave propagation in cylindrical ducts: transmission line parameter approximations for isothermal and nonisothermal boundary conditions. J Acoust Soc Am 1984;75:58–62.

Scoping an acoustic sensing channel for a medical device?

Kyros designs the probe electronics, transducer drive, low-noise capture and firmware for acoustic measurements, and builds the verification evidence that a regulated product needs.

Talk about your project →